General configuration
|
Timeout |
If the default value (10000 ms) is too low or too high (depending on the number of devices at the interface), it can be modified. |
|
Decimal place |
The decimal place for the values shown in the Visu object can be changed here. |
|
WaitTimeHistoryLog |
If the default value (8 s) is too low or too high (depending on the number of devices at the interface), it can be modified. |
Device-specific configuration and parameterization
|
AtmosphericPressure |
The default value (1013 hPa) for the atmospheric pressure can be changed here. Valid value range is from 500 to 1100. |
|
Salinity |
The default value (0.0 g/kg) for the salt content can be changed here. Valid value range is from 0 to 60. |
|
Parameter |
This can be used to switch between three measurands (%Sat, mg/L, and ppm) (display in the Visu object). |
Calibration
Press the button to start calibration.
The measuring range must be selected first.
Slope and two-point calibration can be selected here.
The signal source must now be selected for the temperature here.
Temperature input (sensor signal) or keyboard entry can be selected here.
The air pressure in hPa now has to be input before you can skip to the next step.
The electrode first has to be cleaned here, then immersed into the reference solution. Only when the displayed value is stable can the calibration be started.
In the last step there is an overview of the acquired values and whether the calibration has been performed successfully. The calibration can now be adopted or rejected via the buttons.