General configuration
|
Timeout |
If the default value (10000 ms) is too low or too high (depending on the number of devices at the interface), it can be modified. |
|
Decimal place |
The decimal place for the values shown in the Visu object can be changed here. |
|
WaitTimeHistoryLog |
If the default value (8 s) is too low or too high (depending on the number of devices at the interface), it can be modified. |
Device-specific configuration and parameterization
None available.
Calibration
Press the button to start calibration.
Since there is only zero point calibration, you can immediately skip to the next step.
The actual calibration can now be started.
The value of the buffer solution can be input here and you can skip to the next step.
The electrode first has to be cleaned here, then immersed into the buffer solution. Only when the displayed value is stable can you skip to the next step.
In the last step there is an overview of the acquired values and whether the calibration has been performed successfully. The calibration can now be adopted or rejected via the buttons.