JUMO digiLine O-DO H10/20 (202612)

General configuration

Timeout

If the default value (10000 ms) is too low or too high (depending on the number of devices at the interface), it can be modified.

Decimal place

The decimal place for the values shown in the Visu object can be changed here.

WaitTimeHistoryLog

If the default value (8 s) is too low or too high (depending on the number of devices at the interface), it can be modified.

image-20260714-061518.png

Device-specific configuration and parameterization

SamplingRate

The default value (10 s) for the sampling rate can be changed as desired. Sampling rates from 0 to 9999 seconds are possible.

Calibration

Press the button to start calibration.

image-20260714-063445.png

The calibration method must be selected first.

image-20260714-063457.png

The humid and dry method can be selected here.

image-20260714-063510.png

The calibration procedure must then be selected.

image-20260714-063524.png

Zero point and end value calibration can be selected here.

image-20260714-063538.png

You are now requested to immerse the sensor in the reference solution and to confirm with the forward arrow.

image-20260714-063550.png

Only when the displayed value is stable can you skip to the next step.

image-20260714-063604.png

Input the reference value now and start the calibration.

image-20260714-063618.png

The sensor needs a moment to process the calibration data internally.

image-20260714-072858.png

The result is displayed when processing is complete. The calibration can now be adopted or rejected via the buttons.

image-20260714-072912.png