JUMO ecoLine NTU (202670)

General configuration

Timeout

If the default value (10000 ms) is too low or too high (depending on the number of devices at the interface), it can be modified.

Decimal place

The decimal place for the values shown in the Visu object can be changed here.

WaitTimeHistoryLog

If the default value (8 s) is too low or too high (depending on the number of devices at the interface), it can be modified.

image-20260714-074812.png

Device-specific configuration and parameterization

None available.

Calibration

Press the button to start calibration.

image-20260714-074821.png

The measuring range must be selected first.

image-20260714-074831.png

Four measuring ranges which can be calibrated are available to choose from here.

image-20260714-074842.png

The electrode first has to be cleaned here, then immersed into the reference solution. Only when the displayed value is stable can you skip to the next step.

image-20260714-074901.png

The reference value must then be input. You can only skip to the next step once a valid value has been input.

image-20260714-074911.png

Only when the displayed value is stable can you skip to the next step.

image-20260714-074923.png

The result is displayed when processing is complete. The calibration can now be adopted or rejected via the buttons.

image-20260714-074934.png