Tolerance bands

This part describes the parameters of the tolerance bands.

Parameter

Description

In use

This switch defines the use and the visibility of the tolerance band in JUMO smartWARE Program.

Actual value

This selection defines an actual value that is compared to its setpoint.

Lower deviation min. value

This value defines the smallest possible input value for the lower deviation value.

Entering a permissible deviation of the actual value from 0 in JUMO smartWARE program, deactivates the tolerance band. To avoid this deactivation, the lower limit must be set > 0.

Lower deviation max. value

This value defines the largest possible input value for the lower deviation value.

Upper deviation min. value

This value defines the smallest possible input value for the upper deviation value.

Entering a permissible deviation of the actual value from 0 in JUMO smartWARE program, deactivates the tolerance band. To avoid this deactivation, the lower limit must be set > 0.

Upper deviation max. value

This value defines the largest possible input value for the upper deviation value.

Event type

This selection defines the eventtype behavior in case of an deviation. Possible selections are:

  • None

  • Event

  • Alarm

  • Event and alarm

Tolerance band behavior

This selection defines the program behavior in case of an deviation. Possible selections are:

  • Program stop

  • No influence

Lower deviation default value

Initial setpoint when adding the process step to a program.

Upper deviation default value

Initial setpoint when adding the process step to a program.

The tolerance bands must be configured in the JUMO smartWARE Setup variTRON. They can be found as follows:
Configuration > PT Configuration > Process steps > Process steps > 'Name' > Definition